Nucleus 
Coag-A-Mate MTX II 
Hitachi 747 
Electra 1000 
Elecsys 2010 
Synchron CX9 
Electrolyte 5 Analyzer 
CELL DYN 3200 
CELL DYN 1600 
Axsym 
AVL OMNI 
Stat Profile M 
StaRRsed 
Sysmex 
TEST 1 
UriScan-S300 

 

 


 StaRRsed


1. RS 232C Connection

StaRRsed

Yeungnam Interfacer

25 pin

9 Pin

25 Pin

2

2

2

3

3

3

7

5

7

4, 5 : jump

 

 

2. StaRRsed Configuration

  • Set Baud Rate on the RS232 board.
  • F2 - 261 - E
    • Select RS 232C port parameters.
    • Parameters recommanded
      • Baud ---- 2400,
      • Data Bits  ---- 8,
      • Stop Bits ---- 1,
      • Parity ---- NONE,

3. Yeungnam Interfacer Configuration

  • Download Protocol(YnStaR.zip) and extract zip file.
  • Execute Yeungnam Config.
    • Configure/Protocol Format/Import....  import YnStaR.ynu
    • Configure/Parsing protocol ---- modify protocol for your lab and save it.
    • Other Configure Setting.
  • Execute Yeungnam.

** IF Including raw data in zip file is not compatible to your lab, you can receive raw data from your instruments using "Yeungnam Acquisition", and modify protocol with raw data.

4. Protocol Download