Electrolyte 5 Analyzer, NOVA Biomedical
1. RS 232C Connection
|
Electrolyte 5 |
Yeungnam Interfacer |
|
25
pin |
9
Pin |
25
Pin |
|
2 |
2 |
2 |
|
3 |
3 |
3 |
|
7 |
5 |
7 |
2.
Electrolyte 5 Analyzer Configuration
- Main
Menu-Setup Menu-Devices Setup Menu-Lab Computer
- Set 232C port
parameters.
- Parameters recommanded
- Baud ---- 9600
- Data Bits ---- 8
- Stop Bits ---- 1
- Parity ----
NONE
- Handshake ---- NONE
- Flow Ctrl ---- NONE
- Protocol ----
NONE
- Auto Send ---- ON
- Auto Receive ---- OFF
3.
Yeungnam Interfacer Configuration
- Download Protocol(CRT5.zip) and extract zip
file.
- Execute Yeungnam Config.
- Configure/Protocol Format/Import.... import
YnCRT5.ynu
- Configure/Make or Edit protocol ---- modify protocol for your lab
and save it.
- Other Configure Setting.
- Exit Yeungnam Config.Execute Yeungnam.
** IF
Including raw data in CRT5.zip is not compatible to your lab, you can receive
raw data from your instruments using "Yeungnam Acquisition", and modify protocol
with raw data.
4. Protocol Download
|