[Home
[Yeungnam Interfacer
[Interface protocol
[CELL DYN 3200] 

 

Nucleus 
Coag-A-Mate MTX II 
Hitachi 747 
Electra 1000 
Elecsys 2010 
Synchron CX9 
Electrolyte 5 Analyzer 
CELL DYN 3200 
CELL DYN 1600 
Axsym 
AVL OMNI 
Stat Profile M 
StaRRsed 
Sysmex 
TEST 1 
UriScan-S300 

 

 


 CELL-DYN 3200, Abbott 


1. RS 232C Connection

Cell Dyn 3200

Yeungnam Interfacer

9 pin

9 Pin

25 Pin

2

3

2

3

2

3

5

5

7

 

 2. CELL DYN 3200 Configuration

Main-Setup-Operation Setup-Computer Setup

1.        Auto-transmission of ALTERTED parameter data : ON or OFF

2.        Auto-transmission of NON-ALTERTED parameter data : ON or OFF

3.        Auto-transmission of ALTERTED graph data : OFF

4.        Auto-transmission of NON-ALTERTED graph data : OFF

5.        Transmission CTS enabled : OFF

6.        Set Interface parameters.

  • Parameters recommanded
  • Baud Rate ---- 9600
  • Data Bits  ---- 8
  • Parity ---- None
  • Stop bits ---- 1
  • time out ---- 1

3. Yeungnam Interfacer Configuration

  • Download Protocol(CD3200.zip) and extract zip file.
  • Execute Yeungnam Config.
    • Configure/Protocol Format/Import....  import YnCellDyn32.ynu
    • Configure/Make or Edit protocol ---- modify protocol for your lab and save it.
    • Other Configure Setting.
  • Execute Yeungnam
** IF Including raw data in YnCellDyn32.zip is not compatible to your lab, you can receive raw data from your instruments using "Yeungnam Acquisition", and modify raw data with same procedures.

4. Protocol Download