[Home
[SYNCHRON CX Series] 

 

 

ASASI 
Yeungnam Interfacer 
AU5400 Series 
ADVIA 120 
SYNCHRON CX Series 
Barcode Generator 
Help File 
Table Repair 
Connection with Oracle 
ODBC 
Disclaimer 
Links 

 


Yeungnam Interfacer For CX Series , BECKMAN

(version 0.60, Build 5.)


1. General Information

2. RS 232C Connection

3. CX Configuration

4. Yeungnam Interfacer Configuration

5. Program Download

6. Connection with Oracle 8i


1. General Information

  • I tested this software with SYNCHRON ALX CX9 only. But I think it works with any SYNCHRON CX Series, because CX3, CX4, CX5, CX7, CX4CE, CX5CE, or CX7 system uses same interface protocol.
  • I tested this software for download mode with CX9 at our hospital(Yeungnam University Medical Center), but not query mode. if you have some problems in query mode, please let me know.
  • Including "Host's Test ID", "Instrument's Test ID", "Communication Parameter" , etc are our hospital parameters, you need some modification for your lab.
  • If Including raw data in YnCx060.zip is not compatible to your lab, you can receive raw data(YnCXConfig/Protocol/Receive Raw Data) from your instruments and make parsing protocol suitable to your lab.
  • This version supports "Text File" and "Oracle 8" in host communication.
  • System Requirements
  1. Windows 95 or Windows 98
  2. At least Pentium 200 MHz is recommended.
  3. At least 16Mb of RAM is recommended.
  4. BDE(borland Database Engine)
  5. ODBC Drivers
  6. 1024 x 768 resolution

2. RS 232C Connection

Cx Series

Yeungnam Interfacer

9 Pin (Female)

9 Pin

25 Pin

2

3

2

3

2

3

5

5

7

 

3. CX Series Configuration

    3-1. Host communication parameters setting

Menu: SPECIAL FUNCTION[ F4 ]  in MASTER SCREEN  \  SYSTEM SETUP[ 4 ]  \  HOST COMMUNICATION PARAMETERS[ 10 ]

Example. Our hospital communication parameters

MODE

BIDIRECTIONAL

Baud Rate

9600

Data Bits

8

Stop Bits

1

Parity

NONE

Device ID

0

Flow Control

NONE

Automated Sample Handling

OFF

    

    3-2. Message Option Setting

       Press F1 Key in "Host Communication Parameters SCREEN"

    In Download Mode in our hospital(Batch Download Mode)

    Messages

    Status

    Stream 700/Func 2

    ON

    Stream 702/Func 3

    ON

     

     

    Other Messages  are

    OFF

     

    In Query Mode in our hospital

     

 4. Yeungnam Interfacer Configuration

  • 4.1. Download YnCX06xx.zip, extract zip file and install, See "readme.txt"

    4.2. Execute Yeungnam CX Configure.

    4.3. Other Configure Setting.

    4.4. Exit Yeungnam CX Configure.

    4.5. Execute Yeungnam CX.

     

ver 0.60. Build 2

  • Fix Host's Test ID Setting Error
  • Add Sample Type
  • Fix other bugs
  • add Result Print

     

5. Download